2915-1: Triton Dilution Fridge IOC #2828
Comments
|
The Triton details can be found at : C:\LabVIEW Modules\Drivers\Triton |
Communication : TCP Port : 33576. Make sure the Oxford software is running first. The computer may not respond to a ping due to firewall settings. |
|
Impeded @Tom-Willemsen waiting for labview driver stable |
The command READ:DEV::HTR:SIG:CURR returns INVALID as a command. This is in both open loop and closed loop modes. As a result, I do not believe this approach is possible. I have checked this with the hardware and shown Tom so that he is aware. |
Removed the offending command and replaced it with one that works ( |
As a scientist working on MERLIN, I would like to be able to use a dilution fridge for some very low temperature experiments.
Acceptance Criteria
Notes