LSMAQ

LSMAQ is a lightweight and flexible laser scanning microscope acquisition software written in MATLAB. It supports National Instruments hardware for galvo-based scanning. Present capabilities include:

  • pulse synchronization with MHz-rate lasers (as in 3P microscopy)
  • tiled volume acquisition
  • arbitrary plane 3D scanning
  • phase-stepping for wavefront shaping / deep imaging

Should I use it?

Most likely not. If you are an end user of two-photon microscopes, you will probably be well served by existing software packages like ScanImage. The ScanImage team does a very good job at keeping their software up to date with new developments that most paying end users are interested in.

To minimize code and maximize flexibility, LSMAQ exposes all scanning parameters, performs few checks on user input and generally assumes that the user knows the hardware limitations of their setup. The LSMAQ team cannot offer support to end users.

LSMAQ was created by microscope developers aiming for increased flexibility and ease of customisation. Scripting and quick code modification are facilitated by a clear separation between scanning engine and UI frontend, as well as a lightweight and minimal code base. Adding a new custom property to the UI takes a single line. If you are looking for flexibility or free software for three-photon microscopy, give LSMAQ a try.

Requirements

  • MATLAB (tested with version 2018b, earlier versions since 2009a likely supported). The DAQ toolbox is not required as LSMAQ directly calls NI-DAQmx functions through MATLAB's .NET interface.
  • National Instruments data acquisition hardware: at least one high-speed (MHz) multifunction NI DAQ board. Tested with NI USB-6356 (primary) and NI USB-6343 (secondary)

Installation

  1. Install NI-DAQmx, making sure to include .NET support (tested with version 18.5, earlier versions likely supported)
  2. Clone this repository or copy the contents of this folder to your hard drive.
  3. Review defaultConfig.m to confirm that is reflects your DAQ hardware and wiring (see below)

Wiring

LSMAQ will flexibly support your custom DAQ card wiring as long as that wiring is configured in defaultConfig.m or your custom configuration file. To make your life easier and minimise the edits needed, we suggest using the following default hardware wiring:

Primary DAQ card (e.g. NI USB-6356):

  • PMT inputs: AI ports 0 and 1 (up to 8 input channels supported)
  • Galvo outputs: AO ports 0 and 1
  • Connect PFI12 to USER1 using the screw terminals (USER1 will be the trigger output port)
  • Connect USER1 to PFI0 (trigger input)
  • Connect PFI1 to the shutter
  • Optional: wire P0.0 to USER2 and use as laser blanking signal (to laser or Pockels cell)
  • Optional: connect the laser's pulse sync signal to PFI5 (tested with 1 MHz pulse rate). Adjust sample timing with the cable length of PMT signal and laser sync cables (~5 ns/m coax cable)

Optional: Secondary DAQ card for additional AO channels (e.g. NI USB-6343):

  • Additional outputs (e.g. to Pockels cell or piezo): AO ports 0, 1, etc
  • Connect PFI7 on the primary DAQ card to PFI7 on all other cards (AO clock sync)
  • Connect USER1 on the primary DAQ card to PFI0 on all cards (trigger signal)

Configure / customize

defaultConfig.m

This file defines the startup scan and grab properties, as well as your hardware configuration. Open this file and review the rig configuration section to make sure the settings reflect your hardware and wiring. You can also modify the grab and scan configuration in this file or start LSMAQ (section below) to adjust the properties in the UI.

Properties

Grab properties Description
dirName Save path. Image data and scan configuration are saved as .mat files in HDF5 format. Files can be opened in MATLAB and any other software that supports HDF5 (including FIJI/ImageJ with the HDF5 Plugin)
fileBaseName Each file name will start with this string
fileNumber The next file name will end with this number
nFrames Grab this many frames (per slice/tile)
stackNumXyz 3-element vector. When acquiring slices or tiles, acquire [x y z] slices/tiles along x, y, and z
stackDeltaXyz 3-element vector. Slice/tile separation along [x y z]. Example: to take a stack of 50 slices separated by 5 µm along Z, set stackNumXyz to [1 1 50] and stackDeltaXyz to [0 0 5]. To acquire a tiled volume of two tiles along X, 3 along Y and 50 slices along Z, with an X/Y tile separation of 200 µm, set stackNumXyz to [2 3 50] and stackDeltaXyz to [200 200 5].
stackSequence 3-character string. Aquire slices and tiles in this order. Example: to first move along Z, then X then Y, set this to 'ZXY' (default).
powerDecayLength The power decay length constant in µm. When taking z-stacks, the laser power will be adjusted according to p = p0⋅e(-z/powerDecayLength). To disable power adjustment, set to Inf (default).
Scan properties Description
bidirectional Toggle bi-directional scanning. true or false
fillFraction Fraction of the line not used for flyback. Recommendation for a standard 1.25 MHz analog input rate: 0.8192 (this is 1024/1250; having 1024 fill samples per 1 ms line makes binning by orders of 2 convenient)
sampleLag Lag of the true position versus command position in analog input samples.
nInSamplesPerLine Number of sample per line (including flyback). This setting sets the line rate. At 1.25 MHz, 1250 samples per line would result in a line rate of 1 kHz and 2500 samples in 0.5 kHz.
nLinesPerFrame Number of horizontal lines (vertical image size in pixels)
nPixelsPerLine Number of pixels per horizontal line (horizontal image size). This setting determines sample binning. nPixelsPerLine has to be a divisor of nInSamplesPerLine * fillFraction
piezoStepsN Number of slices that a piezo-based objective scanner is scanning along Z. Default value (without scanner): 1
phaseStepsN Number of phases scanned by a phase-stepper. Phase stepping is performed for each line. When stepping 4 phase steps, the Y mirror will only advance every 4 lines, so that each line is scanned 4 times. phaseStepsN has to be a divisor of nLinesPerFrame. Default value (no phase stepping): 1
scanAmp 2-element vector with X and Y scan amplitudes. Output voltages will be multiplied with this value. The default value, [1 1], will result in a -1V to +1V voltage command signal to both galvos at zoom 1. Directions can be flipped by scaling with negative values. Example: [1 -1] will flip the Y-axis. Optional: 5-element vector with scan amplitudes for X, Y, Z-piezo, Pockels-cell, phase-stepper.
scanOffset 2-element vector with X and Y scan offsets. Output voltages will be offset by this value (after scaling). Default: [0 0]. Optional: 5-element vector with scan offsets for X, Y, Z-piezo, Pockels-cell, phase-stepper.
scanAngle 3-element vector of rotation angles (in degrees) along X, Y, or Z axis. Example: for a 30 deg rotation in the XY plane, set this property to [0 0 30]. With the help of an objective piezo, this can be used for arbitrary plane rotation in 3D, but make sure that the hardware supports the new fast axis direction (or slow down scanning until it does. Use sparingly and keep in mind that only one mirror is designed to be the fast one. Default: [0 0 0]
zoom Image zoom. Higher values result in smaller image sizes. Keep in mind that scanning speed may have to be lowered for lowest zoom levels (highest scan amplitudes). The zoom can also be set by using the mouse scroll wheel in the UI. Example: If scanAmp = [1 1] and zoom = 2, the output voltage range to the X galvo will be -0.5V to +0.5V (the output voltage range to the Y galvo depends on the image size in pixels, as the pixel aspect ratio is kept at 1 when both scanAmp values are the same).

Quick start

Type lsmaq to run. This should open a property window and one or more channel windows.

gui screenshot

Use the property window to edit the configuration. Press enter when done editing or click outside the text field. Start scanning by pressing the play button. Adjust the zoom with the mouse scroll wheel. Stop by pressing play again or by pressing the stop button. Grab with the record button. Record stacks or tiled volumes with the blue button. Close the LSMAQ UI by closing the property window (press 'X').

To control LSMAQ from the command line, use this alternative way to start lsmaq: [rig, prop, hIm] = lsmaq;. This will still open a UI as above, but it will also allow you to do the following:

  • You can adjust scan and grab properties (prop.scancfg and prop.grabcfg) using your own code. For example, type prop.scancfg.zoom = 4 in the matlab prompt. This will immediately adjust the UI. The link is bidirectional. If you change a property in the UI, it will be updated in prop.
  • You can start scanning using command-line. Type data = grabStream(rig, prop, hIm); to grab data using the configuration in prop (this will display data while it is acquired. To run entirely without display, omit the third parameter and type data = grabStream(rig, prop);)

To start LSMAQ with custom hardware and/or scan configuration, create a copy of config/templatecfg.m, rename it (e.g. yourcfg.m) and modify it according to your needs. Then provide the configuration file name (without extension) as an argument: [rig, prop, hIm] = lsmaq('yourcfg')

Example 1: acquire 5 images at zoom levels 1 to 5

Start lsmaq with [rig, prop, hIm] = lsmaq;. You can use the UI to find a sample area of interest. Stop scanning, then execute this code to loop through zoom levels.

prop.grabcfg.nFrames = 1;
for i = 1:5
  prop.scancfg.zoom = i;
  data(:,:,:,:,i) = grabStream(rig, prop, hIm);
end

Under the hood

LSMAQ consists of a command-line scan engine with a UI on top.

  • Scan engine
    • grabStream is the main scan function. Its takes hardware information (initialized by rigClass) plus scan/grab configuration (e.g. from defaultProps) and outputs image data. It talks to makeScanPattern to create analog output signals and raw2pixeldata to process raw analog input into image frames.
  • UI
    • lsmaq starts the UI, consisting of channel figures created by chanfig and a property table of a dynamicshell object. Stacks and tiled volumes are saved with the help of matmap, a convenient way to preallocate large data sets on the disk and save them in compressed HDF5 format.
    • dynamicshell is a class to turn a MATLAB structure into an object that can be passed by reference. Its members can be previewed in a simple property table, forming the main part of the LSMAQ property window.

The following examples will illustrate the purpose of these functions, and how the UI sits on top of the command-line code.

Example 2: command-line acquisition without any graphics

%load default scan and grab configuration as a structure
[prop, rigcfg] = defaultConfig; % or [prop, rigcfg] = yourcfg
%initialize your hardware
rig = rigClass(rigcfg);
%change configuration as needed, for example:
prop.grabcfg.nFrames = 10;
prop.scancfg.nLinesPerFrame = 400;
prop.scancfg.nPixelsPerLine = 400;
%acquire:
data = grabStream(rig, prop);

Example 3: same as above, but with a simple UI for changing properties

Here we will turn the prop structure into a dynamicshell object. This object will still look and feel a lot like a MATLAB structure (you can read and write its fields in the usual way), with two important differences: First, you can pass the data by reference. Second, there is a simple UI to inspect its fields graphically.

%load default scan and grab configuration as a structure
[prop, rigcfg] = defaultConfig; % or [prop, rigcfg] = yourcfg
%initialize your hardware
rig = rigClass(rigcfg);
%convert the structure into a `dynamicshell` object.
prop = dynamicshell(prop)
%inspect the object in a property table
prop.inspect
%add two channel figures
hIm(1) = chanfig(1)
hIm(2) = chanfig(2)
%when done, acquire:
data = grabStream(rig, prop, hIm);

This is, more or less, what lsmaq does – plus colored buttons.

Wish list

  • Stage support: right now only the Sutter MP285 stage is supported. Support for new stages should be easy to add once the need arises.
  • Resonant scanning. Two possible avenues would be to build on top of Vidriotech's code for NI FPGA-based acquisition or interface fast DAQ cards, e.g. from Alazartech (see Scanbox or ScanImage B)

Contributors

Benjamin Judkewitz, Ioannis Papadopoulos, Spencer Smith, Maximilian Hoffmann, Diego Di Battista, Malinda Tantirigama