As soon as a downsampling factor is used, the scope does not trigger properly any more.
Known good configuration: Trigger only on CMD_BEAM_INJECTION, disable trigger on CMD_BEAM_EXTRACTION and CMD_START_ENERGY_RAMP
With downsampling mode, triggers are only received, if no events are disabled (Only CMD_BEAM_EXTRACTION) . As soon as any event is disabled, no data is received at all.
That strange behavior continues even if "downsampling_factor = 1" is picked.
(Doing downsampling in the same way than for dal025, which uses a factor of 8 .. seems to work fine there)
The text was updated successfully, but these errors were encountered:
In GitLab by @alexxcons on Apr 21, 2021, 16:23
As soon as a downsampling factor is used, the scope does not trigger properly any more.
Known good configuration: Trigger only on CMD_BEAM_INJECTION, disable trigger on CMD_BEAM_EXTRACTION and CMD_START_ENERGY_RAMP
With downsampling mode, triggers are only received, if no events are disabled (Only CMD_BEAM_EXTRACTION) . As soon as any event is disabled, no data is received at all.
That strange behavior continues even if "downsampling_factor = 1" is picked.
(Doing downsampling in the same way than for dal025, which uses a factor of 8 .. seems to work fine there)