-
Notifications
You must be signed in to change notification settings - Fork 514
New issue
Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.
By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.
Already on GitHub? Sign in to your account
Boluses deleted from History screen reappear after next bolus (MDT 754 pump) #399
Comments
RPReplay_Final1724935161.mp4 |
Note that in the video above, the 0.6U bolus reappeared, but the 0.18U bolus did not. |
I just reproduced the same thing with my 722 on Trio (TCD). Bolused 0.1 U, deleted from Trio, after next Loop cycle the 0.1 U bolus returned) |
This is a known LoopKit issue. @marionbarker has raised it, I have raised it as well in 2023, when I implemented the swipe-to-delete feature for iAPS that is also found in Trio. It has to the with the event store of the boluses, it is cleared upon running a new cycle. So if you were to bolus 3 times, then delete those boluses, then bolus a 4th time, all within one cycle, all 4 entries will reappear, because they are send to Trio / iAPS / Loop by the device layer event store when the apps perform the storing of pump events to pump history. That is why the 0.18 U bolus did not reappear, but the 0.6 U bolus that was given at the beginning of the recording reappears. We cannot address this in Trio, it needs addressing in LoopKit. Given, that this has not been addressed in ~10 months since finding the issue, I doubt it will ever be addressed. |
Same issue, see previous comment. TCD or non-TCD does not matter here. You can also reproduce this for Loop and for iAPS. A fresh cycle will clear the event store. |
Describe the bug
If you delete a bolus from the history screen, the deleted bolus reappears in the history the next time you issue a bolus.
Attach a Log
Tap the Trio settings icon at the bottom of the screen, then tap 'Share logs' on the bottom of the list and attach it to this ticket.
To Reproduce
Expected behaviour
Once you delete a bolus from the history it should stay deleted. Deleting a bolus from the history may be necessary if your cannula has leaked, or if you have forgotten to suspend while your pump was disconnected.
Screenshots
video attached
Setup Information (please complete the following information):
Smartphone:
Pump:
CGM:
Trio Version:
The text was updated successfully, but these errors were encountered: