Hot Film Sampling with LabJack

This repository contains code to use a Labjack T7 to sample at least 4 hotfilm AD channels at up to 2 KHz and record the channels to disk in NIDAS dat archive format.

Links:

Specific information on the hot films deployed for the M2HATS project can be found on the hot films wiki page on the M2HATS Logbook Wiki.

Running

Systemd service

For the M2HATS project, the service unit has already been installed for the daq user on ustar and t0t. See Install for details.

Starting and stopping with systemctl

systemctl --user start hotfilm
systemctl --user stop hotfilm

Showing run status

Below are commands to check the run status through systemd. See Diagnostics section for ways to monitor the actual data.

(base) [daq@ustar hotfilm]$ systemctl --user status hotfilm.service 
- hotfilm.service - Hotfilm Data Acquisition
     Loaded: loaded (/home/daq/.config/systemd/user/hotfilm.service; linked; preset: disabled)
     Active: active (running) since Sun 2023-07-30 15:34:46 MDT; 1h 34min ago
   Main PID: 1084344 (hotfilm)
      Tasks: 11 (limit: 38243)
     Memory: 235.1M
        CPU: 31.620s
     CGroup: /user.slice/user-1000.slice/user@1000.service/app.slice/hotfilm.service
             └─1084344 /opt/nidas-dev/bin/hotfilm --xml hotfilm.xml

Jul 30 15:34:48 ustar.field.eol.ucar.edu bash[1084344]: 2023-07-30,15:34:48|INFO|Stream started. Actual scan rate: 2000.00 Hz (10000.00 sample rate)
Jul 30 15:34:49 ustar.field.eol.ucar.edu bash[1084344]: 2023-07-30,15:34:49|INFO|creating: raw_data/hotfilm_20230730_213448.dat
Jul 30 15:38:55 ustar.field.eol.ucar.edu bash[1084344]: 2023-07-30,15:38:55|INFO|inet:192.168.1.10:31000: setNonBlocking(true)
Jul 30 16:00:01 ustar.field.eol.ucar.edu bash[1084344]: 2023-07-30,16:00:01|INFO|creating: raw_data/hotfilm_20230730_220000.dat

The log can also be shown with journalctl:

journalctl --user -u hotfilm

Scheduling priority

The hotfilm program tries to set a realtime FIFO scheduling policy with priority 50. It must be given permission for this either by starting as root or by having the appropriate capabilities on the installed file. This is more important on a DSM and less important, if not completely unnecessary, on ustar.

If the file capabilities have been set, then hotfilm can be started as a non-root user but will still be able to set realtime schedule priority:

$ getcap /opt/local/nidas-buster/bin/hotfilm 
/opt/local/nidas-buster/bin/hotfilm cap_net_admin,cap_sys_nice=ep
$ /opt/local/nidas-buster/bin/hotfilm --xml hotfilm.xml 
2023-07-19,00:32:30|INFO|nchannels=4, resolution=4, scanrate=2000, scans_per_read=1000, pps=on, settling=0, range=10
2023-07-19,00:32:30|INFO|thread policy=1 (FIFO=1), priority=50
...

Note the executable file has the cap_net_admin capability, but technically only cap_sys_nice is necessary to set realtime priority. The extra capability prevents some confusing error messages when the NIDAS library tries to add effective capabilities with cap_set_proc(), even though the effective capabilities are already set on the file. See Install.md.

The other option is to start hotfilm as root so it has permissions to set the scheduling parameters, and then switch to a different user with the -u argument. The -E argument to sudo preserves the environment so project environment variables like DATAMNT can be expanded in the XML file:

daq@dsm214:~/hotfilm $ sudo -E -s  hotfilm -u daq --xml hotfilm.xml 
[sudo] password for daq: 
2023-06-29,21:32:57|INFO|thread policy=1, priority=50
...

Configuration

The default settings are 4 channels at 2 KHz at the resolution index 4 (out of 8), with the PPS counter and timestamp synchronization enabled. See hotfilm -h for all the available command-line arguments. Resolution index 4 has a maximum rate of 2.2 KHz for the LabJack T7 for 4 channels. This is based on the table on the T-series data rates page.

Data output

Like for a NIDAS dsm process, the hotfilm process is configured with an XML file which can specify multiple sample outputs, usually a file archive and a sample server socket. The default hotfilm.xml file includes the typical NIDAS output streams, in particular an archive file data stream and real-time sample output on port 31000.

Unlike the sensors in a normal dsm process, the samples recorded by hotfilm are not raw character streams which have to be processed to generate samples with numbers corresponding to variables. Instead, the output samples already contain floating point numbers (float type) broken out into different samples for each channel and different samples for diagnostics. The sample IDs can be used to differentiate and filter them easily.

In hotfilm.xml, the sensor class in the definition is just a placeholder. That sensor class never actually reads any data into samples or processes any raw samples.

At the moment, the IDs in the hotfilm output samples are hardcoded to match the ones in the hotfilm.xml file. When recording samples, the hotfilm.xml file is only used to setup the sample outputs. When reading the output samples, the XML is used to asociate tag information with the samples according to the sample IDs.

Current versions of NIDAS utilities like data_dump and data_stats expect file archives to have only raw samples containing character data. They will filter samples at best or possibly crash if -p is used with the files. This is being fixed on the buster branch and will eventually be fixed in a NIDAS release.

Plotting

There is a python web application script, built on the bokeh framework, for plotting the channel data in real time. It runs data_dump connected to the hotfilm sample output, parses the channel arrays, then updates them in the bokeh app so the web client can plot the updates in real-time. Run the web application like so:

bokeh serve --show app_hotfilm.py --args sock:localhost:31000

The last argument is passed to the data_dump command, so use whatever works to connect to the running hotfilm program. For example, the above command starts the web app on ustar when hotfilm is running on ustar. The --show argument automatically opens a browser window to the local app server instance on the right port.

For the web app to run on ustar but accept connections from browsers on other hosts, use this command on ustar:

bokeh serve --allow-websocket-origin=192.168.1.10:5006 --port 5006 app_hotfilm.py --args sock:localhost:31000

Then browse to url http://192.168.1.10:5006/.

The browser app can plot one of the four channels in either the time or frequency domain, for each second of data output as a sample.

Real-time data can be simulated by passing raw data files as arguments to dump_hotfilm.py, along with the --delay argument to insert pauses between scans:

bokeh serve --show app_hotfilm.py --args hotfilm_20230908_100000.dat --delay 1

This is a typical web plot for M2HATS:

Typical web plot

Diagnostics

This is an example of using data_dump to show all the 1-second statistics and diagnostics without the full 2000-point 1-second time series:

daq@dsm214:~/hotfilm $ data_dump -i -1,501 -i -1,510-513 hotfilm_20230629_203645.dat 
2023-06-29,21:32:08|INFO|opening: hotfilm_20230629_203645.dat
2023-06-29,21:32:08|NOTICE|parsing: hotfilm.xml
|--- date time --------|  deltaT   id          len data...
2023 06 29 20:36:45.7860       0 200, 501      20          1        428          0         48        506 
2023 06 29 20:36:45.7860       0 200, 510      12 0.00052848 -0.0036095  0.0049213 
2023 06 29 20:36:45.7860       0 200, 511      12  0.0046688 0.00081343   0.052004 
2023 06 29 20:36:45.7860       0 200, 512      12 -0.0075408  -0.036148   0.014717 
2023 06 29 20:36:45.7860       0 200, 513      12 0.00053158 -0.0074004  0.0087132 
2023 06 29 20:36:46.7860       1 200, 501      20          2        428          0         96        510 
Exception: EOFException: hotfilm_20230629_203645.dat: open: EOF

Sample 501 has 6 variables:

  • PPS count: latest counter value for the PPS DIO channel. This will be 0 if no PPS pulses have been ever counted by the LabJack. It should normally increment by one for each sample.
  • PPS step: index into the 1-second sample at which the PPS pulse counter changes (steps), from 0 to 1999. If -1, then no change in the PPS counter was detected in this sample, so the time tags were not adjusted to synchronize with the PPS time.
  • Device scan backlog: scans left in the device buffer after the last read; should be near zero and not increasing.
  • Host scan backlog: scans still in the host-side buffer; should be near zero and not increasing.
  • Time of last read in ms: this should be close to 500 ms. The stream is configured to read half a second of samples at a time. So the time spent in the read call should be mostly waiting for a half-second of scans to fill up, or about 500 ms. If it gets small then the host scan reading has fallen behind the LabJackM library buffer and is catching up. If it gets large then the reads from the device are being delayed, such as by network congestion or delays in the LJM library itself.
  • Timetag to system time: This is the difference in microseconds from the sample time to the system time used to derive the sample time. When a PPS counter change is detected in a scan, the sample time is computed according to the PPS step and the current system time. (See Time Tagging.) If the step happens in the second scan of a sample, meaning the PPS step index is 1000-1999, then the offset should be close to 1 second, or 1e6 microseconds. Otherwise, when the step happens in the first scan, then the offset should be close to half a second, or 5e5 microseconds.

Samples 510-513 are the mean/min/max for channels 0-3 over the full second of data in the corresponding samples 520-523.

Showing stats on any of the samples can indicate if there are any synchronization issues. The rate and the min/max time between samples should be 1.0 on any continuously running sample stream:

daq@dsm214:~/hotfilm $ data_stats -i 200,501 hotfilm_20230629_213259.dat  hotfilm_20230629_220000.dat 
2023-06-29,22:38:38|NOTICE|parsing: hotfilm.xml
Exception: EOFException: hotfilm_20230629_220000.dat: open: EOF
sensor  dsm sampid    nsamps |------- start -------|  |------ end -----|    rate minMaxDT(sec) minMaxLen
        200    501      3939 2023 06 29 21:32:59.101  06 29 22:38:37.104    1.00  1.000  1.001   20   20

Logging can also be helpful. Turn on debugging log messages with --log debug. The --diag command-line argument enables extra LJM calls to report on the TCP buffer status and check for skipped scans. However, for normal operations, that probably adds more overhead than it's worth.

Fill values

The LJM stream read function can return data which contains dummy values of -9999. There is documentation (below) with some description of the condition, but it does not say exactly where the -9999 values come from. The LJM_eStreamRead() call can return data with dummy values even without returning an error value, so it seems like the only way to know if scans have been missed is to check if any values are dummy values.

A data_dump of just the 501 samples looks like this when the buffer underrun occurs:

2023 08 08 00:16:02.7705       1      24      22746        459          0         99        508 5.6186e+05
2023 08 08 00:16:04.1315   1.361      24      -9999       1737          0       1402       1201 1.4037e+06
2023 08 08 00:16:04.3500  0.2185      24      22748       1300          0         17        294 1.4802e+06
2023 08 08 00:16:05.7705   1.421      24      22749        459          0         65        519 5.4248e+05
2023 08 08 00:16:06.7705       1      24      22750        459          0         10        446 5.7726e+05

The T-Series Datasheet explains that the LJM library adds the dummy values according to how many scans the Labjack device dropped due to a device buffer overflow. The datasheet seems to be the primary, if not only, reference for the device buffer size setting, STREAM_BUFFER_SIZE_BYTES. The dummy values could happen because the labjack device buffer fills up due to network delays to the host. However, if the device buffer is overflowing, then perhaps the device backlog metric is not useful, since it is always reported as 0.
The LJM library has a buffer size setting named LJM_STREAM_BUFFER_MAX_NUM_SECONDS, which is supposed to default to 20 seconds as long as the system memory supports it. Furthermore, if the LJM library buffer overflows, then the stream read returns LJME_LJM_BUFFER_FULL, so the problem is not in the LJM buffer. There are also timeout configuration settings for the LJM library which affect LJM_eStreamRead(), such as LJM_SEND_RECEIVE_TIMEOUT_MS, but by default the stream read is not supposed to timeout, so that seems an unlikely culprit.

Further development will see if changing STREAM_BUFFER_SIZE_BYTES can reduce the occurrence of dummy values.

Broken hotfilm

If a hotfilm is broken or not installed, then the bridge reports a relatively constant voltage, usually around 2.47 V:

2023 07 21 18:00:10.7670       0 200, 510      12     2.4747     2.4744      2.475
2023 07 21 18:00:10.7670       0 200, 511      12     2.4737     2.4734     2.4741
2023 07 21 18:00:10.7670       0 200, 512      12     2.4969     2.4965     2.4975
2023 07 21 18:00:10.7670       0 200, 513      12     2.4757     2.4753     2.4763

Shorted hotfilm

It has happened at M2HATS that a hotfilm probe mount shorts, so the bridge tries to maintain voltage to heat something which has no resistance, and the LabJack will see a voltage which jumps between 0 and about 7.3 V:

2023 07 22 23:20:43.4780       0 200, 511      12  0.0027043    -0.3827      7.815
2023 07 22 23:20:44.4780       0 200, 511      12     1.6994    0.13827     7.3154

The web plots look like this:

Shorted plots

Shorting cap

A shorting cap can be plugged into the BNC inputs on the LabJack D37 board, in which case the voltage on that channel should be close to zero and steady.

Implementation Notes

The hotfilm program runs similarly to the NIDAS dsm process. However, rather than DSMEngine controlling the sensor opening, polling, and reading, the hotfilm program calls the LabJackM library to open the device and read the stream, blocking where needed. This simplifies the logic of the program and the use of the LabJackM library. The program is also built outside of the NIDAS source tree, so it can be built easily on the DSM3 Pi against buster branch, and so the LabJackM library does not need to be linked into NIDAS.

The LabJackM library does provide a callback API using SetStreamCallback, so the callback can be used to notify when the stream buffer is full and ready to read with LJM_eStreamRead. That could allow the stream reads to be integrated with the NIDAS SensorHandler, if the LabJack sensor provided something like a file descriptor on a pipe to which the callback could write to indicate data are ready to be read.

Stream Mode

This page of the T-series datasheet implies that a 2 KHz scan rate should use Stream mode rather than command-response mode, so that is what hotfilm uses.

PPS Counter

The DIO0 channel is configured as a counter to detect PPS pulses. Given the register settings below:

DIO0_EF_ENABLE=0
DIO0_EF_INDEX=8
DIO0_EF_ENABLE=1

Then DIO0_EF_READ_A will be the current counter, and that channel can be streamed also.

On the DSM3, PPS is on GPIO26. That can be wired to the LabJack FIO0 input using the handy stackable Pi-EzConnect breakout board. One of the GND screw terminals on the breakout has to be wired to the GND input next to the FIO0 terminal.

The LabJack T7 also supports external triggers, so in theory the PPS could be used to trigger the start of a scan, and that might save some overhead from streaming the pulse counter at the same rate as the AIN channels. However, that option was rejected so that sampling can happen even in the absence of the PPS. If the counter synchronization proves reliable, then the full counter stream could be left out of the recorded data to at least avoid that overhead.

Time tagging

If a change in the PPS counter is detected in the scans for that channel, the sample time tags for all the channels are computed relative to that specific scan, called the pps_step in the code and assigned that variable name in the output sample tags. The step index is -1 if a change in the counter was not detected in the last 2000 scans, otherwise it is in the interval 0-1999.

If the counter changed value in the last read buffer, covering the last half-second, then the current system second should be the second at which the pulse happened, as long as the system time is acquired within a half-second after the read returns. To compute the start of the sample, the system time is truncated to the even second, then that time is decremented by the amount of time represented by the step index, pps_step/2000. The sample data are not shifted in any way so that the sample time falls on an even second. Instead the time tag is adjusted so that the time of the sample at pps_step falls on the even second, since that is when the pulse happens.

If there is ever a significant delay, more than half a second, between the last scan in the LabJack buffer and the return of the system time after the buffer is read, then the system time could advance to the next second, past the second corresponding to the PPS pulse in the counter channel. This would be evident by a 2-second difference between successive sample times.

This 0-second or 2-second difference is detected by the time tag algorithm, and the time tag is adjusted to align it with the expected time tag. However, this has some risk. If the very first timestamp is off by a second, then all subsequent timestamps will keep being adjusted and will also be off by one second.

Just in case such a situation ever needs to be detected and corrected, the difference between the timestamp and the system time is stored in the PPS statistics sample as variable timetag_to_system. Adding that value to the sample time tag yields the system time that was returned after the stream read completed. That is the time from which the sample time tag is derived, either by backing off according to pps_step, or else by backing off a full second from the system time after the second read a get an approximate start of the scans in the previous read.

Here are a few thoughts about how to guard against time tag errors from such a significant delay, if it were to become a problem. There could be a check that the system time "seems reasonable" relative to the previous sample time and the value of pps_step. Or, since the time spent in the read is likely to coincide with the time covered by the count, then maybe the average of the system time before and after the stream read would be a better starting point to determine the time at the PPS pulse, perhaps interpolated by the step index so the time after the read returns is more heavily weighted the larger the step index. This seems risky, though, since there is a chance the time before the read could pull the estimated system time of the pulse into the previous second. The interpolated time might have to be incremented by a half-second before being truncated.

It might also help to read the scan buffer more often. The recommendation in the LabJack Stream examples is to read 1/2 the scan rate at a time. However, maybe reading 1/4 at a time makes it more likely to detect the count change in the right second of system time, since there is then 3/4 of a second between the last scan being read and when the system time call must return.

Differential

All the analog inputs are configured as differentials, so each channel requires a pair of analog input terminals. Channel 0 is AIN0+ and AIN1-, channel 1 is AIN2+ and AIN3-. In the code, the even inputs are named in the scan list: AIN0, AIN2, AIN4, AIN6.

Resolution

The scan resolution is set to 4 by default, the highest possible resolution index for the 2 KHz scan rate and 4 channels.

LJM provides the option to read 16-bit data instead of converting to float on the host side. However, there does not seem to be any disadvantage to recording the data already scaled to Volts. One downside is that 32-bit floats take twice as much space, but that is unlikely to be a problem.

Interference

When testing the LabJack in the lab, the power source for the LabJack seems to affect the digitization when also using a waveform generator. When powered through a laptop, the digitized waveform signal looks clean. When powered by the DSM USB or by the LabJack AC-USB adapter, the signal drops towards 0 V at somewhat regular intervals. Here is an example:

Power source interference.

The interference does not seem to happen without the waveform generator connected, regardless of the LabJack power source. Or so we hope.

4 KHz Sampling

See Sampling4K.md for notes on the switch from 2K to 4K sampling.

Example Data

sawtooth.dat

  • channel 0: AA battery reading ~1.7V
  • channel 1: sawtooth waveform 1 Vpp and 1 V offset, 50 Hz, but actually reading at 2V offset and 1 Vpp

LabJack was powered by the USB port on the waveform generator.

sinewave.dat

  • channel 0: hotwire through the bridge
  • channel 1: 990 Hz 5 mVpp sine wave

The hot wire was blown on several times and then fanned it with a flat box lid.

Todo

The diagnostic samples can be viewed in real-time using the tips above, but they are not yet available through a DSM dashboard. The diagnostics could be relayed as UDP packets to another DSM instance, which then includes those variables in the dashboard. Or hotfilm could run on its own DSM with its own dashboard, and the json_data_stats service could be configured to dump only the diagnostic samples.