-
Notifications
You must be signed in to change notification settings - Fork 514
New issue
Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.
By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.
Already on GitHub? Sign in to your account
[alpha-issue] - TidePool - Looks like all boluses are labeled as "automated". This highlighted bolus was manual. #77
Comments
@MikePlante1 I think oiaps doesn't store the origin (SMB or manual bolus) of each dose injected. Probably requires to improve the data model of DoseEntry. |
@dnzxy , I believe the distinction between manual bolus and SMB was introduced after v2.3.3, perhaps you were involved in that? |
Yes. See #93, @BrianWieder is working in bringing that over. I implemented it for iAPS, Brian ported it over (thanks bud!). It sets a flag, so you should be able to differentiate between SMBs, manual doses and external doses @avouspierre |
Thanks! I do see the |
I think Brian will bring that in judging by #93 description.
It was part of Artificial-Pancreas/iAPS#267 and later automatic was removed from History view and replaced with SMB 😊 |
Yes, going to work on bringing over isSMB and displaying it within the history sheet later today. |
Created the |
I believe this issue is resolved now? Which PR fixed this? |
@bjornoleh the correction was done with the PR #104 |
Hi, PR #104 was actually closed and not merged into dev. Perhaps it was replaced with another PR, or already merged in alpha? I could however not find anythingh in Trio dev when searching fro Could you please check if the fix for automated vs manual bolus in Tidepool is indeed included in current dev? |
hey 👋 - silence for 30 days 🤐 ... anybody? triage is required! |
I think this is resolved, but maybe not by the PR that was referenced above. |
This is not resolved. I just tested it. All boluses are labeled automatic in the Tidepool website display:
In Loop the distinction is made that manual and external (non-pump) both show up as manual (no label shown and a lighter blue) and automatic shows up as automatic (label says automatic and darker blue). |
hey 👋 - silence for 30 days 🤐 ... anybody? triage is required! |
This issue is still present in main branch. |
Addressed for TCD / Trio 1.0.0 dev. |
(This test was done via the alpha branch, not specifically this PR, btw)
Originally posted by @MikePlante1 in #30 (comment)
The text was updated successfully, but these errors were encountered: