- Light artifacts are dramatically reduced by ramping light instead of square-wave light pulses (1 ms ramp time is sufficient, this requires a DAQ samplerate of ~40 kHz)
- Orienting the electrode sites TOWARDS the light (circuit boards facing down, because the probe is mounted upside-down relative to the headstage) reduces the light artifact