You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
Problem
Currently recOrder acquires data in the z-first order and toggles the shutter between each frame (if auto-shutter is chosen). The z-first order increases the chances of misalignment between polarization channels when there is drift in stage or when the live sample is being imaged. Opening/closing the shutter can slow down the acquisition of dynamics.
Proposed solution
It is better to set the MDA in channel-first mode and keep the shutter open when label-free channels are acquired.
Additional context
Noted during AQLM 2022
The text was updated successfully, but these errors were encountered:
Problem
Currently recOrder acquires data in the z-first order and toggles the shutter between each frame (if auto-shutter is chosen). The z-first order increases the chances of misalignment between polarization channels when there is drift in stage or when the live sample is being imaged. Opening/closing the shutter can slow down the acquisition of dynamics.
Proposed solution
It is better to set the MDA in channel-first mode and keep the shutter open when label-free channels are acquired.
Additional context
Noted during AQLM 2022
The text was updated successfully, but these errors were encountered: