Anton Paar L‐Dens 3300 Density Sensor
George Ryall edited this page Jan 8, 2025
·
6 revisions
- This device has no commands and asynchronously sends readings.
- Which readings it sends depends on which application is loaded onto the device, at ISIS we use the
Live Raw Data
application at the moment with room to expand, meaning it is safe to assume the device will send data in the following formatT= 24.74 <°C> D= 0.8 <kg/m³><\r><\n>
; a temperature of the fluid and the density of the fluid in degrees Celsius and kilograms per metre cubed respectively. - If a different application is loaded then the temperature and density PVs sit in
INVALID
alarm. - To account for the device's asynchronous behaviour we use I/O interrupt AI records which trigger when the input it is pointing to changes.
- If 10 seconds pass and no messages are received in the expected format then the temperature and density PVs go into
INVALID
alarm.
The IOC to facilitate use of this was implemented under https://github.com/ISISComputingGroup/IBEX/issues/8413.
Baud = 9600
Data = 8
Parity = None
Stop = 1
The manual can be found at \\ISIS\Shares\ISIS_Experiment_Controls\Manuals\AntonParr_L-Dens3300
. (With the operator and admin pass codes on page 16)