Notifications
You must be signed in to change notification settings - Fork 2
INTER Instrument Details
This page collects information that will be useful for the implementation of the IBEX control system on INTER.
INTER is a high-intensity chemical interfaces reflectometer instrument at ISIS, on TS2. The INTER web page describes the background to the instrument.
As of 23/06/2021 INTER is running IBEX.
The equipment listed below is used on INTER. Please add new information (e.g. new items of equipment, new notes, information about drivers, etc. to this table).
Manufacturer | Model | Type | Connection | Driver | Notes |
---|---|---|---|---|---|
ISIS | DAE 2 | Detector Electronics | Ethernet | see DAE note | |
ISIS | Mk3 | Chopper | N/A | see Chopper note | |
GALIL | DMC2280 | Motion Controller | Ethernet | EPICS | |
??? | 4-blade jaws | see Jaws note | |||
ISIS | ??? | ISIS Vacuum System | see Vacuum System note | ||
Pfeiffer | TPG300 | Vacuum Gauge | RS-232 | EPICS | see Vacuum System note |
Eurotherm | Temperature Controller | All models at ISIS | RS-232 | EPICS | see Eurotherm note |
Julabo | FP-50 | Water Bath | RS-232 | see Water Bath note | |
Julabo | FP-52 | Water Bath | RS-232 | see Water Bath note | |
KEPCO | BOP 100-10MG | Bi-Polar Power Supply | RS232 | #187 | see Kepco note |
Keithley | 2410 | Source Meter | RS-232 | see Keithley note | |
Linkam | T95 | RS232 | see Linkam note | ||
Knauer | K6 | Electric valve drive | RS232 | see Knauer note | |
Knauer | 1050 | HPLC pump | ??? | see Knauer HPLC note | |
see Hitachi note | |||||
Jasco | PU-4180 | HPLC pump | ??? | see JASCO HPLC note | |
Keyence | LK-G | Positioning Sensor | RS-232 | see Keyence note | |
Nima Trough | ??? | Trough | ??? | see Nima Trough note | |
Thurlby | EX355P | PSU | ??? | see Thurlby note | |
Omega | iBTHX | Transmitter | Ethernet | see Omega note | |
Sensirion | SHT75 | Humidity Sensor | ??? | see Sensirion note | |
Watson Marlow | 323 | Peristaltic Pump | ??? | see Peristaltic Pumps note | |
WPI | Aladdin-1000 | Syringe Pump | ??? | see Syringe Pumps note | |
WPI | SP2xx | Syringe Pump | ??? | see Syringe Pumps note | |
Biologic | SP200 | Potentiostat | ??? | see Biologic note | |
Weeder | WTDAC-M | Analog Output Card | ??? | see Weeder note | |
ISIS | Environment Monitor | ??? | see ISIS Environment Monitor note |
See multi-detector and single-detector below.
INTER has a Mk3 chopper.
Provide information about INTER jaws. (After reflectometry meeting these seem to be micro jaws which they borrow from another instrument this is not a galil controlled but have not yet got any extra information)
TPG300 support is implemented via #216 and #2063
Eurotherms are used to control temperature Orange Cryostat, CCR and Furnace devices.
INTER uses Julabo water baths
- Julabo (which model(s)?)
-
Kepco BOP 100-10MG PSU.
See also #3005
-
Keithley 2400 Series Source Meter.
See also tickets #1826, #2695, #2801 and #3176.
Hitachi L-7100 HPLC pump (no longer supported at ISIS). Can't find L-7100 on [Hitachi web-site](http://www.hitachi-hightech.com/global/about/corporate/group/hhs (may be obsolete). Check existing SECI VI for logic and manual.
There are VIs for this.
See also Linkam T95 controller and tickets #1106, #1496, #1509.
Knauer K-6 Electric Valve Drive. Knauer web site
K-6 model appears to have been superseded. Check existing SECI VI for logic and manual.
- Support now implemented (28-03-2019). See #3781.
Knauer HPLC 1050 Knauer HPLC 1050 is discontinued.. Check existing SECI VI for logic and manual.
- Support now implemented (04-01-2019). See #3262.
JASCO HPLC pump is a new (for ISIS) model of HPLC pump.
- See
C:\LabVIEW Modules\Drivers\Jasco PU-4180 HPLC Pump\Documentation
for documentation. - See also #3743 & #3923
- Keyence Web-site
- The Keyence LK-G is a laser positioning sensor (possibly superseded by a newer model). It is used when adjusting the height of the sample stage. Check the existing VI.
- See
C:\LabVIEW Modules\Instruments\INTER\Keyence LK-G
for existing VI. - Current models of Keyence LK-G laser sensors
- The Keyence LK-G is always in position on the instrument, although not always used.
- Now implemented (24-07-2019). See #3745
- See
Nima Trough: SECI uses a manufacturer supplied VI (see C:\LabVIEW Modules\Drivers\Nima Trough
). We may need to do the same in IBEX (via lvDCOM).
-
Note: NIMA Technologies Ltd now seems to be part of Biolin Scientific.
- The NIMA trough is used regularly on SURF. The manufacturer supplied VI is used to view graphs showing information about thin films.
- The manufacturer has made additional software available for download. A copy of this software is located in
\\isis\shares\ISIS_Experiment_Controls\NIMA Trough\Nima_TR8.1.zip
. - Support now implemented (25-05-2019). See #3783
Thurlby EX355P PSU - see #155 and #198.
- Thurlby Thandar Instruments EX355P PSU
- IOC & OPI updated (18-07-2019). See #3784
OMEGA™ iBTHX transmitter is a device to monitor and record barometric pressure, temperature, relative humidity, and dew point over an Ethernet network or the Internet.
- OMEGA devices have proved unreliable on other instruments (e.g. ZOOM, LOQ). Consider other types of device.
Sensirion SHT75 is a device to monitor and record humidity, temperature,… etc.
- According to the manufacturer's web-site the SHT7x series of devices is "end-of-life" as of 31/12/2018.
Biologic SP200 is a potentiostat.
- See also #618
-
Watson Marlow 323 Peristaltic Pump
- Support now implemented (14-08-2019). See #3786
-
WPI Aladdin-1000 Syringe Pump
- Support now implemented (20-06-2019). See #3787
-
WPI SP2xx Syringe Pump - check specific model.
- Support now implemented (05-07-2018). See #3261
- Weeder WTDAC-M analog output card.
- The ISIS Environment Monitor is a device to monitor various environmental properties (temperature, pressure, humidity, etc). There is an existing list of commands and a VI to work from.
- Suspect this is the Sensirion SHT75 device.
- Vacuum chamber – specialist setup using HV and vacuum with some safety trips integrated (instrument scientists says this is be a high priority to enable in IBEX).
- Check SECI VI.
- Is the vacuum chamber related to the INTER High Voltage VI (listed below)?
INTER is due to get new mirror guides (prior to Cycle 2019/01). These will be operational (under SECI) before INTER is migrated to IBEX. There is a sketch showing the three mirror guide sections (MG1 - MG3). MG1 is fixed; it requires no control. MG2 and MG3 are moveable and, therefore, do require control.
There will be, essentially, 3 modes of operation:
- no bounce: all three sections centred in height around the incident beam height to allow straight through beam on sample
- one bounce: this could be either from the middle section (MG2) or the focusing section (MG3), the latter being more likely
- two bounces: MG2 and MG3 both intercept the (reflected beam) with their inner bottom surface and deflect the beam up
In modes 2 and 3 several beamline components need to track: intermediate slit (S1b), slit 2, sample monitor, sample height, slit 3 and detector.
In May 2019, SECI was updated to handle the new mirror guides. IBEX will need to replicate the new functionality in SECI. The is a document describing the SECI changes.
Document information about INTER SECI configs here.
Configuration Name | Sub-Configurations | Last Accessed | Required |
---|---|---|---|
INTER_place_holder1.conf | - | dd/mm/yyyy | - |
INTER_place_holder2.conf | - | dd/mm/yyyy | - |
Similarly, Document information about INTER SECI Genie scripts here.
INTER has the following specialist panels:
- INTER Front Panel
- INTER KEYENCE LK-G
- INTER Galil DIO
- INTER High Voltage
- OFFSPEC Piezo Stage
- SURF Galil DIO
INTER has the following devices under motion control:
- FOM
- Jaws & Height
- Jaws
- Laser Gimbal
- Reflectometer Sample Stack
- Transmission Monitor
- INTER Multi Detector
- INTER Point Detectors
- INTER Supermirror